logo Michigan Scientific Corporation

About Us | Contact Us | Site Map | Links | Help

AMP-TC-M1 Miniature Thermocouple Amplifier Specifications

PARAMETER

SPECIFICATION

OUTUT

Output is not linearized over temperature, if necessary, the output may be linearized externally

Range

± 10 V Max

Sensitivity @ 25°C TC Temperature

10 mV/°C

Capacitive load

1000 pF Max

TEMPERATURE ERROR

Includes errors due to cold junction compensator

Initial @ 25°C Case Temperature

± 2°C Max

-25°C to +85°C Case Temperature

± 3°C Max

-55°C to +125°C Case Temperature

± 5°C Max

NOISE

 

0.01 - 10 Hz

0.8 µV p-p

DYNAMIC RESPONSE

 

Frequency Response

1.16 kHz

Slew rate

0.4 V / µs

Settling Time 0.1% / 0.01%

40 µs / 50 µs

POWER REQUIREMENTS

Voltage

± 15 VDC

Quiescent Current

± 5 mA Max Total ( 1 channel )
 
± 10 mA Max Total ( 2 channels )
 
± 15 mA Max Total ( 3 channels )

ENVIRONMENT

Specification

-25 to +85°C ( -13 to +185°F)

Operation

-55 to +125°C ( -67 to +257°F)
Version: 3.0 | ©2005 Michigan Scientific Corporation | Updated: May 14, 2008 11:04 AM