logo Michigan Scientific Corporation

About Us | Contact Us | Site Map | Links | Help | Chinese

Model AMPEH2-HS Twelve Channel Strain Gage Spinning Amplifier Specifications

PARAMETER

SPECIFICATION

BRIDGE EXCITATION

Customer specified excitation level

Type

Constant Voltage

Magnitude

5,10, or 17 V

Current Limit

100 mA

Temperature Coefficient

0.0005%/°C Max (0.00028 %/°F)

REMOTE CALIBRATION

Positive & negative solid state shunt calibration

Shunt value range

100 k ohm or greater

Shunt accuracy

0.1% or better

GAIN

Customer specified

Range

1 to 10,000 V/V

Accuracy @ 25°C, Gain =500

± 2%

Temperature Coefficient

0.0014%/ °C

OUTPUT

 

Range

±10 V

Capacitive Load

1000 pF Max

VOLTAGE OFFSET*

Referred to input of amplifier

Initial** @ 25°C, Gain = 500

±15 µV

Temperature Stability @ Gain = 500

± 0.2 µV /°C

Time Stability @ Gain = 500

±1.0 µV / Month

INPUT VOLTAGE

 

Range

Varies with configuration

CMRR @ DC

128 dB

CMRR @ 60 Hz / 100 Hz

110 dB @ 100 Hz

NOISE

Referred to input of amplifier

0.01 - 10 Hz

0.8 µV p-p

DC - 1 Hz

--

DYNAMIC RESPONSE

 

Frequency Response ± 3 dB / 1%

10 kHz / 1 kHz @ Gain 100

Slew rate

0.6 V/ µs

Settling Time 0.1%/0.01%@Gain=100

80 µs / 150 µs

POWER REQUIREMENTS

 

Voltage @ 25°C

±13 to ±17 VDC

Current

±15 mA plus Bridge Load (+15 mA additional during shunt calibration)

POWER REQUIREMENTS

 

Voltage

± 15 VDC

Current

± 30 mA plus Bridge Load

ENVIRONMENT

 

Specification

-25 to +85°C (-13 to +185°F)

Operation

-55 to +125°C (-67 to +257°F)

Storage

-65 to +150°C (-85 to +302°F)

* Upon bridge excitation voltage

**Average (channel to channel)

Version: 3.0 | ©2005 Michigan Scientific Corporation | Updated: October 17, 2008 4:10 PM